Q
How do I calibrate a microscope using an ocular micrometer and stage micrometer?
▼
A
Place the stage micrometer on the microscope stage and focus on its scale at the magnification you wish to calibrate. Align the ocular micrometer scale with the stage micrometer divisions and count how many ocular divisions correspond to a known stage micrometer distance. Divide the known distance by the number of ocular divisions to determine the value per ocular division at that magnification.
Q
What is an ocular micrometer and how is it installed?
▼
A
An ocular micrometer is a glass disc with an engraved scale that is placed inside the eyepiece of a microscope. Unscrew the eye lens from the eyepiece, place the micrometer disc on the field-stop diaphragm inside the eyepiece (scale facing down), then re-screw the eye lens. The scale appears superimposed on the sample image.
Q
How often should microscope calibration be performed?
▼
A
Calibration should be performed whenever the objective or eyepiece is changed, when the tube length is adjusted, and at regular intervals (typically annually) for quality assurance. If measurement results seem inconsistent, recalibration is recommended.
Q
Do I need a different calibration for each objective magnification?
▼
A
Yes. The value of each ocular micrometer division changes with objective magnification. Calibration must be performed separately for each objective (e.g., 4×, 10×, 40×, 100×) and the results recorded. When switching objectives, use the calibration value corresponding to that magnification.